LSP01(04)-1A Syringe Pump Calibration

Release Time:2017-01-06

The flowrate can be calibrated for more accurate dispense volume.

1. In parameters setting interface, press and turn RES to highlight the Calibration line. Press RES to enter calibration interface.

2. Testing time line is highlighted. Press RES to highlight the numerical value position. Turn RES to select the testing time. The testing time is from 0.5 minutes to 60 hours. Press RES to confirm.

3. Press Run/Pause key to start the testing. After testing, turn RES to highlight the testing volume line. Press RES to highlight the numerical value position. Enter the actual volume from the membrane keypad. Press RES to confirm.

4. After calibration, the linear speed of the Pusher Block changes to increase the accuracy of the dispense volume.