WT600-1F - Dispensing Peristaltic Pump Calibration

Release Time:2016-12-23

Pump should be calibrated when flow rates or dispensing volume is out of the tolerance. Flow rates and dispensing volume can be calibrated separately in different applications. 
*Flow rates calibration
1. Press and turn Knob to highlight calibration line. Press Knob to enter next interface. Turn Knob to select Flowrate. Press Knob to highlight test time line. Press Knob again to highlight the value of time. Turn Knob to select test time (its range is from 30 to 1800 seconds), press Knob for confirmation. Press Exit Key to return to previous menu.
2. Press Start/Stop Key, the LCD displays running time and actual volume. Running time counts down.
3. Turn Knob to input actual test volume. Press Knob for confirmation. The calibration operation can be repeated.

If the actual volume is known in advance, enter the calibration interface and input the actual volume directly.

*Dispensing volume calibration: 
The parameters can not be changed under dispensing volume calibration operation. The defaulted parameters are the parameters set under Dispensing Setting.
1. Pump stops under dispense mode, press and turn Knob to highlight calibration line. Press Knob to enter next interface. Turn Knob to select Dispense.
2. Press Start/Stop Key, the LCD displays running time and actual volume. Running time counts down.
3. Turn Knob to input actual volume. Press Knob for confirmation.

If the actual volume is known in advance, enter the calibration interface and input the actual volume directly.