How to use External Control Function for BT600-2J

Release Time:2016-12-23



Analog Control Mode
External control module is shown as below. When using external control function just trigger the External Control Switch in external control position and insert external control module to the external control interface which is in the rear panel of peristaltic pump, and then connect the corresponding control signal wire.


Terminal Definition of Standard External Control Module
1# Analog speed control input, for external speed control in external control modules of 4 - 20 mA, 0 - 5V and 0 - 10 V.
2# External start/stop control input. When connected to COM,the pump runs, when connected to high level (5 - 12 V), the pump stops.
3# External cw/ccw control input. When connected to COM, the pump rotates cw, when connected to high level (5 - 12 V), the pump rotates ccw.
4# Reference potential for 4 - 20 mA, 0 - 5V/10V speed control input. Pulse speed control input, for external speed control in external control module of 0 - 10 KHz.
5# Reference potential for external cw/ccw, external start/stop and pulse speed control input.